• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number ATG120144
Medical Device Problem Codes Deflation Problem (1149); Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/07/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an angioplasty procedure via left popliteal vein, the pta balloon was allegedly unable to aspirate.Despite repeated aspirations, the pta balloon was unable to deflate.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, video was provided for review.The investigation of the reported event is currently underway.(expiry date: 03/2023).
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one atlas gold pta dilatation catheter with unknown inflation device was returned for evaluation, inflation catheter was cut, and segment was returned attached to inflation luer of the hub was returned for evaluation.No specific anomalies noted during the visual evaluation.During the functional testing, the balloon was inflated to 6 atm by inhouse presto inflation device.The balloon inflated and maintained pressure.Balloon took two minute and twenty seconds to deflate.The balloon was cut and under the microscopic observation it was noted that both port holes were deformed.No further functional testing was performed.The deflation time is not within the acceptable range as per the product specification limit.One video was reviewed.The video shows a fluoroscopic image of a partially inflated angioplasty balloon with a stent.The exact location of this is likely within the right iliac vasculature.Therefore, based on the video review, the reported deflation issue could not be confirmed on the findings no conclusion can be made.One video was reviewed.The balloon was noted to be in inflated condition connected with an unknown inflation device, the plunger within the device was noted to been in the initial position.No evidence of deflating the balloon could not be noted.Therefore, based on the video review, the reported deflation issue could not be confirmed on the findings no conclusion can be made.He investigation for the deflation issue was confirmed as the balloon time took two minute and twenty seconds to deflate under laboratory conditions and also the deflation was not within the acceptable range of product specification limit.During the microscopic observation it was noted that both port holes were deformed which may lead to the reported deflation issue.However, definitive root cause for the reported deflation issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4(expiry date: 03/2023).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event or Problem Description
It was reported that during an angioplasty procedure via left popliteal vein, the pta balloon was allegedly unable to aspirate residual amount of contrast agent and it was further reported that the pta balloon was allegedly slow to deflate.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens 75751
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13152947
Report Number2020394-2021-02145
Device Sequence Number5348033
Product Code DQY
UDI-Device Identifier00801741060762
UDI-Public(01)00801741060762
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K181323
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberATG120144
Device Catalogue NumberATG120144
Device Lot Number93PE0118
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/07/2021
Supplement Date Received by Manufacturer02/10/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date02/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age53 YR
Patient SexMale
Patient Weight62 KG
-
-