• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE PS FEM LT SZ 3 CEM; ATTUNE IMPLANT : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE PS FEM LT SZ 3 CEM; ATTUNE IMPLANT : KNEE FEMORAL Back to Search Results
Model Number 1504-10-103
Medical Device Problem Codes Tear, Rip or Hole in Device Packaging (2385); Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/12/2021
Type of Reportable Event Malfunction
Event or Problem Description
Inner packaging damaged and open while the outer packaging is sealed.Surgical delay of 3 minutes occurred.
 
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> the device associated with this report was returned.Visual examination of the provided photographic evidence and the returned product denoted that the inner and outer blister of the packaging were melted/deformed.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot
=
> ¿product code 150410103 work order (b)(4) was manufactured on 16 july 2021.¿ 12 parts were manufactured to specification.¿ there was no scrap associated with this lot.¿ 01 non-conformances associated with this lot: 1.(b)(6) is related to a digital callipers wall thickness (snap gauge) failed calibration.There is no correlation between this non-conformance and the failure mode of the complaint.¿ there were no deviations associated with this lot.Device history review
=
> a manufacturing record evaluation was performed for the finished device lot/serial/batch number, and no related non-conformances / manufacturing irregularities were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FEM LT SZ 3 CEM
Common Device Name
ATTUNE IMPLANT : KNEE FEMORAL
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13153589
Report Number1818910-2022-00088
Device Sequence Number5348626
Product Code NJL
UDI-Device Identifier10603295041603
UDI-Public10603295041603
Combination Product (Y/N)N
Initial Reporter CountryTC
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1504-10-103
Device Catalogue Number150410103
Device Lot Number9800986
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2021
Supplement Date Received by Manufacturer05/11/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date05/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-