| Model Number |
1504-10-103 |
| Medical Device Problem Codes |
Tear, Rip or Hole in Device Packaging (2385); Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Inner packaging damaged and open while the outer packaging is sealed.Surgical delay of 3 minutes occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary = > the device associated with this report was returned.Visual examination of the provided photographic evidence and the returned product denoted that the inner and outer blister of the packaging were melted/deformed.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot = > ¿product code 150410103 work order (b)(4) was manufactured on 16 july 2021.¿ 12 parts were manufactured to specification.¿ there was no scrap associated with this lot.¿ 01 non-conformances associated with this lot: 1.(b)(6) is related to a digital callipers wall thickness (snap gauge) failed calibration.There is no correlation between this non-conformance and the failure mode of the complaint.¿ there were no deviations associated with this lot.Device history review = > a manufacturing record evaluation was performed for the finished device lot/serial/batch number, and no related non-conformances / manufacturing irregularities were identified.
|
| |
|
Search Alerts/Recalls
|