ARTHREX, INC. IBAL UKA TIB BEA-RING TRL SZ 5 9MM; TEMPLATE FOR CLINICAL USE Back to Search Results
Model Number IBAL UKA TIB BEA-RING TRL SZ 5 9MM
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On (b)(6) 2021, it was reported by a sales representative via e-mail that an ar-601-tbe9 tibial insert trial broke upon removal from the knee.This was discovered during an uka procedure on (b)(6) 2021.The broken fragment was retrieved and the case was completed.
 
Additional Manufacturer Narrative
Complaint confirmed.One unpackaged ar-601-tbe9 was received for investigation.Visual inspection identified that the trial body had broken along the outer diameter, with surface damage present across the remaining material.The thickness of the returned ar-601-tbe9 was assessed using digital micrometer id: 224, and was identified to meet design specifications.A probable cause is attributed to damage incurred during the removal process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IBAL UKA TIB BEA-RING TRL SZ 5 9MM
Common Device Name
TEMPLATE FOR CLINICAL USE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key13154403
Report Number1220246-2022-04237
Device Sequence Number14764418
Product Code HWT
UDI-Device Identifier00888867037694
UDI-Public00888867037694
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 01/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIBAL UKA TIB BEA-RING TRL SZ 5 9MM
Device Catalogue NumberAR-601-TBE9
Device Lot Number5261640
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2021
Supplement Date Received by Manufacturer12/13/2021
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date01/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/19/2016
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-