MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMPLATZ SUPER STIFF; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMPLATZ SUPER STIFF; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 39575
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the guidewire distal tip detached.A.035 x 260 cm amplatz guidewire was selected for use in a filter retrieval.During preparation, while the device was external to the patient, the distal tip was observed to be detached.
 
Event or Problem Description
It was reported that the guidewire distal tip detached.A.035 x 260 cm amplatz guidewire was selected for use in a filter retrieval.During preparation, while the device was external to the patient, the distal tip was observed to be detached.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the device was returned for analysis.Unit returned with its original pouch, overall visual revision did not identify failures or evidence that could be lost due to decontamination process.The returned device was inspected and it had the distal tip kinked.It's important to mention that the device was received outside the original pouch and protective hoop.No more visual damages were found in the device.Detailed pictures of the tip damage were taken and it was possible to observed the distal tip kinked.The outer diameter of the distal tip, middle section and proximal section were within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZ SUPER STIFF
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13155271
Report Number2134265-2021-16285
Device Sequence Number14489856
Product Code DQX
UDI-Device Identifier08714729295747
UDI-Public08714729295747
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K843015
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 02/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/07/2024
Device Model Number39575
Device Catalogue Number39575
Device Lot Number0028169759
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/10/2021
Supplement Date Received by Manufacturer01/10/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date02/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age62 YR
Patient SexMale
-
-