| Catalog Number |
H93821 |
| Medical Device Problem Code |
Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the packaging of a fluid transfer tube set was damaged.The event was further described as; uncovered (open) device is detected, without the security seal.This issue was identified before patient use.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter phone no.: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
H4: the lot was manufactured from september 23, 2020 - september 29, 2020.H10: the device was received for evaluation.Unaided visual inspection was performed which observed that the bottom side of the primary packaging was opened and not properly sealed.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|