• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE : SYNTHES USA PRODUCTS LLC GUIDE BLOCK FOR 2 COLUMN PLATE 6 HOLE HEAD/LEFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE : SYNTHES USA PRODUCTS LLC GUIDE BLOCK FOR 2 COLUMN PLATE 6 HOLE HEAD/LEFT Back to Search Results
Catalog Number 03.111.601
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/07/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional narrative: 510k: this report is for an unk - guides/ sleeves/aiming/ unknown lot.Part and lot numbers are unknown; udi number is unknown.Complainant part is not expected to be returned for manufacturer review/ investigation.Without a lot number, the device history records review could not be completed as no product was received.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, the patient underwent open reduction internal fixation surgery for fracture of distal radius.In the surgery, the screw of the guiding block broke.The broken part of the screw stuck to the hole of the plate and could not be removed.The surgeon gave up using the plate and used another new plate.The surgery was completed with no surgical delay.No further information is available.This complaint involves two(2) devices.This report is for (1) unk - guides/ sleeves/aiming.This report is 2 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).510k: this report is for an unk - guides/ sleeves/aiming/ unknown lot.Part and lot numbers are unknown; udi number is unknown.Complainant part is not expected to be returned for manufacturer review/ investigation.Without a lot number, the device history records review could not be completed as no product was received.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, the patient underwent open reduction internal fixation surgery for fracture of distal radius.In the surgery, the screw of the guiding block broke.The broken part of the screw stuck to the hole of the plate and could not be removed.The surgeon gave up using the plate and used another new plate.The surgery was completed with no surgical delay.No further information is available.This complaint involves two(2) devices.This report is for (1) unk - guides/ sleeves/aiming.This report is 2 of 2 for (b)(4).
 
Event or Problem Description
This report is for one (1) guide block for 2 column plate 6 hole head/left.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: a product investigation was conducted.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the guidingblock f/2-column dist-rad-pl2.4 6 (product code: 03.111.601 lot number: 10-3114) is broken from the fixing screw.A dimensional inspection was not performed for the guidingblock f/2-column dist-rad-pl2.4 6 (product code: 03.111.601 lot number: 10-3114) due to post manufacturing damage.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for guidingblock f/2-column dist-rad-pl2.4 6 (product code: 03.111.601 lot number: 10-3114).There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h4, h6: a device history record (dhr) review was conducted: part # 03.111.601, lot # 10-3114, manufacturing site: werk selzach, supplier: (b)(4), release to warehouse date: 08 jul 2010.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDE BLOCK FOR 2 COLUMN PLATE 6 HOLE HEAD/LEFT
Common Device Name
GUIDE
Manufacturer (Section D)
WRIGHTS LANE : SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach 2545
SZ   2545
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13155483
Report Number8030965-2022-00044
Device Sequence Number17495693
Product Code FZX
UDI-Device Identifier07611819352537
UDI-Public(01)07611819352537
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.111.601
Device Lot Number10-3114
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/07/2021
Supplement Date Received by Manufacturer02/15/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date03/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/08/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VA-LCP-2-COLUMN DRP2.4 VOLAR NARR LE SHA
Patient SexUnknown
-
-