| Model Number |
1012450-12 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a mildly calcified, mildly tortuous first obtuse marginal artery that is 90% stenosed.The 3.25x12mm nc trek balloon dilatation catheter (bdc) had been fully deflated, but was stuck with the 014 hi-torque turntrac flex guide wire as bdc was moved.The guide wire and the balloon were removed together as one unit.Another nc trek was used with a non-abbott guide wire to successfully complete the procedure.There was no adverse patient effects and no clinically significant delay.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.The hi-torque turntrac flex guide wire referenced is filed under a separate medwatch report number.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported complaint.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Search Alerts/Recalls
|