• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 15750 ALTON PKWY RAD-8, HORIZONTAL; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 15750 ALTON PKWY RAD-8, HORIZONTAL; OXIMETER Back to Search Results
Model Number 22042
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2021
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported unit does not accurately provide measurements.No consequences or impact to patient were reported.
 
Additional Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Additional Manufacturer Narrative
Additional manufacturing narrative: the returned device was evaluated.During evaluation the device passed all visual and functional testing.During simulation testing, the device passed manual and preset conditions and provided accurate measurements.The unit was found to visually and audibly alarm during alarm conditions.The unit was determined to be functioning as designed.
 
Event or Problem Description
The customer reported unit does not accurately provide measurements.No consequences or impact to patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-8, HORIZONTAL
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - 52 DISCOVERY
52 discovery
irvine CA 92618
Manufacturer Contact
matofa fautua
15750 alton pkwy
irvine, CA 92618
9498223661
MDR Report Key13155633
Report Number3019388613-2022-00003
Device Sequence Number14476927
Product Code DQA
UDI-Device Identifier00843997005647
UDI-Public00843997005647
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K053269
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 12/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number22042
Device Catalogue Number9190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2022
Supplement Date Received by Manufacturer01/10/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date02/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/08/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-