ABBOTT VASCULAR 0.014 HI-TORQUE (HT) TURNTRAC GUIDE WIRE Back to Search Results
Model Number 1020009
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.The nc trek balloon is filed under a separate medwatch report number.
 
Event or Problem Description
It was reported that the procedure was to treat a mildly calcified, mildly tortuous first obtuse marginal artery that is 90% stenosed.The 3.25x12mm nc trek balloon dilatation catheter (bdc) had been fully deflated, but was stuck with the 014 hi-torque turntrac flex guide wire as bdc was moved.The guide wire and the balloon were removed together as one unit.Another nc trek was used with a non-abbott guide wire to successfully complete the procedure.There was no adverse patient effects and no clinically significant delay.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.Factors that may contribute to the difficult to remove the guide wire from the balloon dilatation catheter (bdc) causing resistance between the devices may include, but not limited to, manufacturing damage, inner diameter of guide wire lumen, condition of the guide wire, or damage to the distal shaft of the catheter.The product risk assessment identifies this as a foreseeable event; as such, design and manufacturing controls and mitigations have been established for potential causes.The investigation was unable to determine a conclusive cause for the reported complaint.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
0.014 HI-TORQUE (HT) TURNTRAC GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13155691
Report Number2024168-2022-00032
Device Sequence Number12132344
Product Code DQX
UDI-Device Identifier08717648222627
UDI-Public08717648222627
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K173795
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2022
Device Model Number1020009
Device Catalogue Number1020009
Device Lot Number0022761
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/11/2021
Supplement Date Received by Manufacturer01/20/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date02/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/27/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
3.25X12MM NC TREK
Patient SexUnknown
-
-