• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON, S.A. BD PHASEAL¿ PROTECTOR P50; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON, S.A. BD PHASEAL¿ PROTECTOR P50; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 515105
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/06/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd phaseal¿ protector p50 experienced leakage between the protector and vial.The following information was provided by the initial reporter: the customer had a leakage with the protector.
 
Additional Manufacturer Narrative
H.6.Investigation: no photos or physical samples showing the reported problem were provided for investigation.A device history review was performed for the reported lot 2110120 no deviations or non-conformances were identified during the manufacturing process that could have contributed to this issue.Five retained samples from the same lot were evaluated.The protectors were successfully connected to the vials, ensuring that they fitted correctly.The product was evaluated, and no damage or defects were observed.Functional tests were performed on the samples and no leakage was identified between the protector and the vial.Based on the available information, we are unable to identify a root cause related to the manufacturing process at this time.
 
Event or Problem Description
It was reported that the bd phaseal¿ protector p50 experienced leakage between the protector and vial.The following information was provided by the initial reporter: the customer had a leakage with the protector.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHASEAL¿ PROTECTOR P50
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer (Section G)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13155711
Report Number3003152976-2021-00875
Device Sequence Number5335221
Product Code ONB
UDI-Device Identifier30382905151058
UDI-Public30382905151058
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K123213
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/30/2026
Device Model Number515105
Device Catalogue Number515105
Device Lot Number2110120
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2021
Supplement Date Received by Manufacturer02/09/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date02/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-