• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR PROSTAR XL SUTURE-MEDIATED CLOSURE; VESSEL CLOSURE SUTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR PROSTAR XL SUTURE-MEDIATED CLOSURE; VESSEL CLOSURE SUTURE Back to Search Results
Model Number 12322-01
Medical Device Problem Codes Product Quality Problem (1506); Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The customer reported the device is not returning.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.The additional prostar xl device referenced is filed under a separate medwatch report number.
 
Event or Problem Description
It was reported that an arteriotomy closure of the right common femoral artery was attempted using two prostar xl devices with a 10f sheath after an abdominal aortic aneurysm (aaa) procedure.Reportedly, the needles did not come out through the hub for both devices and deployment felt different.A third prostar xl device was used and felt normal.The third device achieved hemostasis.There was no adverse patient sequela and no reported clinically significant delay in the procedure or therapy.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the manufacturing records identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation was unable to determine a conclusive cause for the reported difficulties; however, the treatment appears to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROSTAR XL SUTURE-MEDIATED CLOSURE
Common Device Name
VESSEL CLOSURE SUTURE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13156249
Report Number2024168-2022-00036
Device Sequence Number17180504
Product Code MGB
UDI-Device Identifier08717648000621
UDI-Public08717648000621
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P960043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Model Number12322-01
Device Catalogue Number12322-01
Device Lot Number1042941
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/14/2021
Supplement Date Received by Manufacturer02/07/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date02/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HEPARIN.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-