U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL LONG NAIL KIT R1.5, TI, LEFT GAMMA3® Ø11X320MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL LONG NAIL KIT R1.5, TI, LEFT GAMMA3® Ø11X320MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 3525-1320S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Implant Pain (4561)
Date of Event 12/04/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Upon completion of investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
As reported: "broken gamma nail, 2 months post-surgery.Need a second medical procedure.Revision of gamma nail + u-blade column screw." update: when walking from toilet to the kitchen the patient suffered acute severe pain.She had to sit down and could no longer load the leg.
 
Event or Problem Description
As reported: "broken gamma nail, 2 months post-surgery.Need a second medical procedure.Revision of gamma nail + u-blade column screw." update: when walking from toilet to the kitchen the patient suffered acute severe pain.She had to sit down and could no longer load the leg.
 
Additional Manufacturer Narrative
The reported event could be confirmed.Review of dhr, complaint history, capa databases and risk analysis did not identify any discrepancies.No nc/capa had been filed for the item in question.Potential nail breakage had been considered in the risk management file.Potential reasons for nail breakage are listed in the labelling.Physical investigation of the nail revealed it broke in a fatigue manner, originated by material damage at lateral in the posterior web of the proximal drill hole.Additionally, the nail was significantly loaded during the implantation period.Material and manufacturing deficiencies were not detected.Provided medical information was reviewed by a hcp who found surgical factors (suboptimal fracture reduction, suboptimal positioning of the lag screw) and patient related factors (fracture pattern, presumably poor bone metabolism because of the high age with reduced fracture healing potential, limited musculotendinous equipment due to the preexisting coxarthrosis) which may have contributed to the nail breakage.From technical point of view the event was caused by intraoperative material damage contributed by increased weight bearing.With available information a deficiency of the device was not verified.The cause of the event was determined being intraoperative material damage contributed by the patient.In case further substantive information becomes available we reserve the right to reopen the investigation with root cause assessment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONG NAIL KIT R1.5, TI, LEFT GAMMA3® Ø11X320MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
marilyne chaumont
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13156511
Report Number0009610622-2022-00004
Device Sequence Number4118189
Product Code HSB
UDI-Device Identifier07613252274261
UDI-Public07613252274261
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K034002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/28/2025
Device Model Number3525-1320S
Device Catalogue Number35251320S
Device Lot NumberK04FE17
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/09/2021
Supplement Date Received by Manufacturer03/22/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date04/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age86 YR
Patient SexFemale
Patient Weight60 KG
-
-