Model Number 1867
Medical Device Problem Codes Entrapment of Device (1212); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/23/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a guidezilla was stuck in the lesion.The target lesion was located in the left circumflex artery.A guidezilla guide extension catheter was selected for use.During the procedure, it was noted that the device was scratched and stuck in the lesion.The device was removed normally with the ancillary device.The procedure was completed with another of same device.No intervention was done and no patient complications were reported.
 
Event or Problem Description
It was reported that a guidezilla was stuck in the lesion.The target lesion was located in the left circumflex artery.A guidezilla guide extension catheter was selected for use.During the procedure, it was noted that the device was scratched and stuck in the lesion.The device was removed normally with the ancillary device.The procedure was completed with another of same device.No intervention was done and no patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.The device was visually and microscopically examined.There was blood present in the shaft of the device.There were numerous shaft kinks.For a length of 1.1cm from the tip of the device, the shaft, marker band, and tip of the device was flattened.Within the length of the flat shaft, there was drag.Product analysis for the reported event of the device being stuck in the lesion could not be performed as the clinical circumstances could not be replicated.However, the shaft kinks, flattened shaft, marker band, and tip found to the device during analysis supports the reported event as such which would indicate device resistance and procedural difficulties.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDEZILLA
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13157049
Report Number2134265-2021-16311
Device Sequence Number4147318
Product Code DQY
UDI-Device Identifier08714729839859
UDI-Public08714729839859
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K123765
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/29/2022
Device Model Number1867
Device Catalogue Number1867
Device Lot Number0026273011
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/24/2021
Supplement Date Received by Manufacturer02/11/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date02/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexMale
Patient Weight73 KG
-
-