• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH STRAIGHT REDUCTION CLAMP, BROAD VARIAX2

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH STRAIGHT REDUCTION CLAMP, BROAD VARIAX2 Back to Search Results
Model Number 705297
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
As reported: "the tip of the product was damaged due to a strong pull when it was used during restoration." update 12/28/2021: the tip of instrument is broken.Other instruments were used to complete the procedure.
 
Additional Manufacturer Narrative
Upon completion of investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
As reported: "the tip of the product was damaged due to a strong pull when it was used during restoration." **update (b)(6) 2021: the tip of instrument is broken.Other instruments were used to complete the procedure.
 
Additional Manufacturer Narrative
Please note correction to section h6 (clinical signs and health impact codes).The reported event could be confirmed, since the device was returned for evaluation, and matches the alleged failure.The device inspection revealed the following: the received clamp overall showed normal signs of usage with one of its legs broken.The fracture pattern on the surface of the broken leg indicates a typical brittle fracture under immense bending load.No other deformation was observed around the breakage point, indicating an instantaneous breakage under a high load.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of manufacturing or design-related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the above investigation, the root cause of the failure is user related.The breakage of one of the legs of the clamp occurred due to intra-op excessive bending load application on the leg, more than it could sustain.Ultimately leading to an instantaneous brittle fracture.If any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRAIGHT REDUCTION CLAMP, BROAD VARIAX2
Common Device Name
CLAMP
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
marilyne chaumont
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13157823
Report Number0008031020-2022-00001
Device Sequence Number4146282
Product Code HXD
UDI-Device Identifier07613327125252
UDI-Public07613327125252
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number705297
Device Catalogue Number705297
Device Lot NumberL28818
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/09/2021
Supplement Date Received by Manufacturer03/09/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date04/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/23/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-