• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH LOCKING SCREW VARIAX 2 T10, FULL THREAD, 3.5MM / L10MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH LOCKING SCREW VARIAX 2 T10, FULL THREAD, 3.5MM / L10MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 657310S
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/10/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that intraoperatively, the locking did not lock, so it was replaced with a new screw.
 
Additional Manufacturer Narrative
Upon completion of investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
It was reported that intraoperatively, the locking did not lock, so it was replaced with a new screw.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the locking thread is damaged.The returned locking screw was inspected under a microscope.The thread below the screw head which is intended to lock the screw in the plate, shows plastic deformation which could explain the loss of the angular locking ability.The damage pattern suggests high force application during locking attempts.A potential non-conformity report was initiated to address this event.The comprehensive root cause analysis of the potential ncr included a surface and microstructure analysis as well as nano-indent hardness measurement of variax1 and variax2 screw samples by the fraunhofer institute on behalf of stryker.Further internal investigation efforts were focusing on the anodization color difference between variax1 and variax2 screws.Significant differences between the screws which could result in the reported complaints have not been revealed in either fraunhofer or stryker investigations.Excessive in-house handling and bench testing with variax2 screws from various manufacturing batches showed that locking is achieved as intended.The desired increase in torque indicating to the customer that the screw is locking in the plate, can be confirmed.A review of the device history for the reported lot did not indicate any abnormalities.No deviation from the specifications was noted.No indications of material, manufacturing or design related problems were found during the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW VARIAX 2 T10, FULL THREAD, 3.5MM / L10MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
sharon rivas
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13158084
Report Number0008031020-2022-00004
Device Sequence Number14447709
Product Code HWC
UDI-Device Identifier07613327086799
UDI-Public07613327086799
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K132502
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Model Number657310S
Device Catalogue Number657310S
Device Lot NumberG31133
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/10/2021
Supplement Date Received by Manufacturer01/19/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date02/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age56 YR
Patient SexFemale
Patient Weight52 KG
-
-