• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS BD SAFETYGLIDE¿ INSULIN SYRINGE WITH ATTACHED NEEDLE; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS BD SAFETYGLIDE¿ INSULIN SYRINGE WITH ATTACHED NEEDLE; PISTON SYRINGE Back to Search Results
Model Number 305903
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/06/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medical device expiration date: unknown.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that the rubber stopper was separated from the plunger while using bd safetyglide¿ insulin syringe with attached needle.There was no report of patient impact.The following information was provided by the initial reporter: "the needle core rubber stopper is separated from the syringe barrel.".
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 2022-01-11.Investigation summary: one photo and one 1ml syringe were received from an unknown batch # (p/n 305903) and evaluated.The photo was of a loose 1ml syringe next to a plunger rod without a stopper.The stopper was inside the syringe bottomed out.There was also an activated safety glide needle attached to the syringe.There was no shield on the needle.The loose 1ml syringe received had an activated safety glide needle attached without a shield.Also, it had a stopper inside the syringe bottomed out.There were two vertical stress marks in the syringe observed.Upon disassembly it was observed that the stopper was jammed inverted inside the syringe.The condition observed was non-conforming per product specification.Potential root cause for the jammed stopper defect, plunger separation and barrel damage is associated with the assembly process.The lot number is unknown, therefore device history record review (dhr) or quality notification review (qn) and corrective action could not be performed.H3 other text : see h10.
 
Event or Problem Description
It was reported that the rubber stopper was separated from the plunger while using bd safetyglide¿ insulin syringe with attached needle.There was no report of patient impact.The following information was provided by the initial reporter: "the needle core rubber stopper is separated from the syringe barrel.".
 
Additional Manufacturer Narrative
Becton dickinson and company's (bd) medication delivery solutions (mds) business unit will discontinue malfunction mdr reporting for certain device failures that have not caused or contributed to deaths or serious injuries in the past two years, and where the likelihood of a death or serious injury as a result of these malfunctions is remote.This decision follows fda guidelines (medical device reporting for manufacturers guidance for industry and food and drug administration staff issued nov 8, 2016, reference section 2.15).Bd notified fda of this decision on mar 3, 2025.Fda has reviewed and responded to bd¿s notification (reference fda document # (b)(4)) on march 7, 2025.This documentation is available in bd document management system (sap) as dir 10000690801.This supplemental mdr is being filed to document that the below device failure will no longer be considered a reportable malfunction mdr for the product family below: product family: safetyglide.Device failure: stopper separation.
 
Event or Problem Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SAFETYGLIDE¿ INSULIN SYRINGE WITH ATTACHED NEEDLE
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13158214
Report Number1213809-2021-00833
Device Sequence Number19870422
Product Code FMF
UDI-Device Identifier30382903059035
UDI-Public30382903059035
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K951254
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number305903
Device Catalogue Number305903
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2021
Supplement Date Received by Manufacturer01/24/2022
Not provided
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date02/07/2022
04/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1