• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC. SECURMARK; MARKER, RADIOGRAPHIC, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC. SECURMARK; MARKER, RADIOGRAPHIC, IMPLANTABLE Back to Search Results
Model Number SMARK-M-SS1
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/14/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a eviva procedure, the marker migrated resulting in the physician having to use ultrasound to place a second marker for surgery.No other information is available.
 
Additional Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SECURMARK
Common Device Name
MARKER, RADIOGRAPHIC, IMPLANTABLE
Manufacturer (Section D)
HOLOGIC, INC.
250 campus drive
marlborough, MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough, MA 01752
Manufacturer Contact
daniel guevara
562 parkway
coyol free zone building b24
san jose, alajuela 20102- CRI
CS   20102 CRI
MDR Report Key13158242
Report Number1222780-2022-00003
Device Sequence Number5478966
Product Code NEU
UDI-Device Identifier15420045503977
UDI-Public(01)15420045503977(17)210323(10)20C23RC
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K062528
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/23/2021
Device Model NumberSMARK-M-SS1
Device Catalogue NumberSMARK-M-SS1
Device Lot Number20C23RC
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/17/2021
Initial Report FDA Received Date01/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-