• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LOGIC CR TIB INSERT SLOPE+, SZ 3, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC CR TIB INSERT SLOPE+, SZ 3, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Model Number LOGIC CR TIB INSERT SLOPE+, SZ 3, 9MM
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/16/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: logic cr femoral cem, left, sz 3 (cat# 02-010-03-0230 / serial# (b)(4)).Lgc tibial fit tray cem sz 3f / 3t (cat# 02-012-45-3030 / serial# (b)(4)).Three peg patella 35mm (cat# 200-02-35 / serial#(b)(4)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported, a (b)(6) female patient had a revision due to effected poly causing poly wear within 15 months.Patient's liner was replaced.There was no breakage of the device.There was no surgical delay or prolongation.Patient was last known to be in stable condition following the event.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections d4, d9, g3, g6, h2, h3, h4 and h6 have been updated accordingly.(h3) the revision reported may have been the result of prosthesis wear as stated in the experience report.Additionally, a contributing factor to the early onset of wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.The most probable root cause associated with the reported event of "prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC CR TIB INSERT SLOPE+, SZ 3, 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key13159742
Report Number1038671-2022-00003
Device Sequence Number17167627
Product Code JWH
UDI-Device Identifier10885862174383
UDI-Public10885862174383
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K111400
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/30/2021
Device Model NumberLOGIC CR TIB INSERT SLOPE+, SZ 3, 9MM
Device Catalogue Number02-012-48-3009
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2021
Supplement Date Received by Manufacturer04/26/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/07/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
Patient SexFemale
Patient Weight63 KG
-
-