Catalog Number 366560
Medical Device Problem Codes Short Fill (1575); Volume Accuracy Problem (1675)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/07/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported when using the bd vacutainer® buffered sodium citrate (9nc) blood collection tubes, the device experienced underfill or low draw of a tube with blood.The following information was provided by the initial reporter.The customer stated: it was reported that once in the vein, blood begins to flow but stops abruptly.We are currently using your 2.7 ml sodium citrate for our nih-funded clinical trial, ignite.We've had significant issues recently collecting blood in these tubes specifically.Once in the vein, blood begins to flow but stops abruptly.At times we have tried up to 8 different tubes within a pack and rarely get one that works.All of bd tubes work seamlessly with the same stick, so we believe there is an issue with the tube.
 
Event or Problem Description
It was reported when using the bd vacutainer® buffered sodium citrate (9nc) blood collection tubes, the device experienced underfill or low draw of a tube with blood.The following information was provided by the initial reporter.The customer stated: it was reported that once in the vein, blood begins to flow but stops abruptly.We are currently using your 2.7 ml sodium citrate for our nih-funded clinical trial, ignite.We've had significant issues recently collecting blood in these tubes specifically.Once in the vein, blood begins to flow but stops abruptly.At times we have tried up to 8 different tubes within a pack and rarely get one that works.All of bd tubes work seamlessly with the same stick, so we believe there is an issue with the tube.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 12/20/2021.H.6.Investigation: bd received 3 samples from the customer in support of this complaint.The 3 sample tubes, along with 10 retention samples of the bd inventory, were functionally tested and the customer's indicated failure mode of underfill was not observed as all tubes were within specification limits.Bd was unable to confirm and/or duplicate the customer¿s indicated failure mode because the defect was not observed in the sample and retention testing.The device history records were reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.H3 other text : see h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® BUFFERED SODIUM CITRATE (9NC) BLOOD COLLECTION TUBE
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13159935
Report Number1917413-2021-01100
Device Sequence Number3978805
Product Code JKA
UDI-Device Identifier00382903665600
UDI-Public00382903665600
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K013971
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date05/31/2022
Device Catalogue Number366560
Device Lot Number1228002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2021
Supplement Date Received by Manufacturer01/05/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date01/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1