• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number MC2-4590S
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Approximate implant date reported as 2019 and approximate explant date reported as (b)(6) 2021.The device has been received and is pending evaluation.The root cause is unable to be determined at this time.A supplemental report will be submitted following the device evaluation.
 
Event or Problem Description
Information was received that the rod failed to reach terminal distraction lengthen.The rods were removed for final fusion with a competitor product.No patient adverse event was reported.
 
Additional Manufacturer Narrative
Additional data d4 (lot number).Evaluation summary: visual inspection was conducted of the returned rod, which revealed the distraction rod of the magec 2 rod had visible score marks.X-ray images of the internal components showed no obvious damage and revealed the rod was partially distracted.The rod was functionally tested and was able to distract and retract with the erc and manual distractor.The minimum and maximum lengths were measured and the total stroke measurement of 48.06 was found to meet specification.Distraction force testing was measured and also found to meet specification.A review of manufacturing records found the rod was manufactured to specified requirements and met all of the required quality inspection release criteria.
 
Event or Problem Description
No additional information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS
101 enterprise #100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS
101 enterprise #100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise #100
aliso viejo, CA 92656
8583448112
MDR Report Key13162870
Report Number3006179046-2022-00104
Device Sequence Number5335460
Product Code PGN
UDI-Device Identifier00812258022624
UDI-Public812258022624
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K171791
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC2-4590S
Device Lot Number8082718AAA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2021
Supplement Date Received by Manufacturer01/20/2022
Initial Report FDA Received Date01/05/2022
Supplement Report FDA Received Date01/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/27/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-