• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. BIOPSY VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. BIOPSY VALVE Back to Search Results
Model Number MB-358
Medical Device Problem Codes Break (1069); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2021
Type of Reportable Event Malfunction
Event or Problem Description
Olympus medical systems corp.(omsc) was informed that during the first use in unspecified procedure, the slit was damaged and the patient's body fluids or filth leaked from air/water valve, suction valve or the biopsy valve.The user replaced it with another device but same issue occurred.Since there was no problem with the procedure, the user used it and completed the procedure.There was no report of patient injury associated with the event.
 
Additional Manufacturer Narrative
The subject device was not returned to olympus for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since it is an accessory.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, it is likely the slit was damaged due to external factors and the liquid leaked.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOPSY VALVE
Common Device Name
BIOPSY VALVE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13163423
Report Number8010047-2022-00865
Device Sequence Number11786024
Product Code FDS
UDI-Device Identifier04953170063947
UDI-Public04953170063947
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K051645
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMB-358
Device Lot Number19A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2021
Supplement Date Received by Manufacturer06/13/2022
Initial Report FDA Received Date01/05/2022
Supplement Report FDA Received Date06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-