• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number ZCU300
Medical Device Problem Code Inaccurate Delivery (2339)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient age and date of birth, patient weight, ethnicity and race: unknown as information was asked but not provided.Date implanted: not applicable as the iol was not implanted.Date explanted: not applicable as the iol was not implanted.The device was not returned for analysis therefore, a visual analysis of the complaint device cannot be completed.A review of the device/lot history record and historical data analysis for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch report will be filed.Attempts were made to contact the customer account requesting additional information regarding complaint however, to date no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported the intraocular lens (iol) misfired and there was patient contact however, extent is unknown.It was confirmed the issue was not due to use error.The reporter believes there was no medical intervention and no surgical intervention.No further information is available.
 
Additional Manufacturer Narrative
Product investigation was received, and the following fields were updated accordingly: section d-9: device available for evaluation: yes.Section d-9: date returned to manufacturer: 14-jan-2022.Section h-3: device evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed viscoelastic residue on the optic body and haptics.A haptic was stuck to the optic body which can be attributed to handling during loading and deployment.The lens was cleaned, and no defects were observed afterwards.The complaint issue could not be confirmed, and no product deficiency could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13165823
Report Number3012236936-2022-00030
Device Sequence Number12170490
Product Code HQL
UDI-Device Identifier05050474660090
UDI-Public(01)05050474660090(17)250617
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/17/2025
Device Model NumberZCU300
Device Catalogue NumberZCU300U170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2021
Supplement Date Received by Manufacturer02/18/2022
Initial Report FDA Received Date01/05/2022
Supplement Report FDA Received Date02/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-