• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON FIBER OPTIC ILLUMINATOR; LIGHT, SURGICAL, FIBEROPTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON FIBER OPTIC ILLUMINATOR; LIGHT, SURGICAL, FIBEROPTIC Back to Search Results
Catalog Number 8065751577
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported the shaft part of the trocar cannula disintegrated and went missing while removing the trocar from the eye during a surgery.The product was replaced and the surgery was completed.There was no patient harm.
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.10.One opened trocar hub in a container was received.The returned trocar hub was visually inspected and was found to be non-conforming, the trocar cannula is missing from the hub.There was presence of adhesive inside of the hub.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.During assembly of the trocar product adhesive is dispensed at the cannula hub interface.The evaluation of the returned sample identified adhesive inside of the hub, therefore, how and when the trocar can during surgery cannula and hub became separated cannot be determined from this evaluation and the root cause for this complaint is unknown.A potential contributing factor of the separation is manipulation the exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.Assemblies are 100% inspected for adhesive application during assembly.If adhesive application is incorrect the assembly is scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBER OPTIC ILLUMINATOR
Common Device Name
LIGHT, SURGICAL, FIBEROPTIC
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13167284
Report Number1644019-2022-00019
Device Sequence Number5478673
Product Code FST
UDI-Device Identifier00380657515776
UDI-Public00380657515776
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K875005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 03/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2023
Device Catalogue Number8065751577
Device Lot Number2456388H
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/14/2021
Supplement Date Received by Manufacturer03/13/2022
Initial Report FDA Received Date01/05/2022
Supplement Report FDA Received Date03/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CONSTELLATION VISION SYSTEM; CONSTELLATION VISION SYSTEM
Patient SexUnknown
-
-