| Catalog Number |
8065751577 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported the shaft part of the trocar cannula disintegrated and went missing while removing the trocar from the eye during a surgery.The product was replaced and the surgery was completed.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in d.9., h.3., h.6.And h.10.One opened trocar hub in a container was received.The returned trocar hub was visually inspected and was found to be non-conforming, the trocar cannula is missing from the hub.There was presence of adhesive inside of the hub.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.During assembly of the trocar product adhesive is dispensed at the cannula hub interface.The evaluation of the returned sample identified adhesive inside of the hub, therefore, how and when the trocar can during surgery cannula and hub became separated cannot be determined from this evaluation and the root cause for this complaint is unknown.A potential contributing factor of the separation is manipulation the exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.Assemblies are 100% inspected for adhesive application during assembly.If adhesive application is incorrect the assembly is scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|