|
SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR ANTI-TG (ATG); SYSTEM, TEST, THYROID AUTOANTIBODY
|
Back to Search Results |
|
| Model Number |
N/A |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/24/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Discordant, low advia centaur xp anti-thyroglobulin (atg) assay results were observed for two patient samples in a method comparison.The initial results from the advia centaur xp analyzer were reported to the physician, and were not questioned.However, repeat testing of the samples by alternate methods produced clinically contradictory results; the repeat results were observed in the context of method-comparison testing, and were not reported.There are no known reports of patient intervention or adverse health consequences due to the discordant atg results.
|
| |
|
Additional Manufacturer Narrative
|
|
A customer from outside the united states observed advia centaur xp anti-thyroglobulin (atg) patient results which were discordant relative to results obtained using alternate methods.Siemens is investigating.The product's instructions for use (ifu) states the following, under interpretation of results: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings." mdr 1219913-2022-00001 was submitted for the same event.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens healthcare diagnostics filed mdr 1219913-2022-00002 on 2022-01-05.Additional information, 2022-02-10: siemens has concluded the investigation.A customer from outside the united states observed advia centaur xp anti-thyroglobulin (atg) patient results which were discordant relative to results obtained using alternate methods.Discordance was identified when comparing results among three assays: advia centaur atg, advia centaur atgii, and immulite 2000 atg.Siemens notes that patient results may vary among these assays for a number of reasons, including differences in assay architecture, standardization, and diagnostic cutoff, as well as the variability in the effect of anti-tg antibodies.[antibodies to thyroglobulin are frequently present in thyroid cancer patients, and can interfere with measurements to differing extents.].No product problem was identified.The customer is operational, and no further investigation is required.Notes: in section h6, the codes for type of investigation, investigation findings and investigation conclusion have been updated on the basis of the investigation results.Mdr 1219913-2022-00001 was submitted for the same event.
|
| |
|
Search Alerts/Recalls
|
|
|