Model Number 364487
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2021
Type of Reportable Event Malfunction
Event or Problem Description
An orsiro drug eluting stent system was selected for treatment of a severely calcified lesion (90 percent stenosis degree).The orsiro stent system could not cross the lesion in the severe tortuous diagonal.Another stent was used to complete the intervention.
 
Additional Manufacturer Narrative
The returned device was subjected to a detailed technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.In addition, the angiographic material provided was reviewed.The technical investigation revealed that the stent is deformed at its distal end, i.E.Few struts are bent.Microscopic inspection showed stent imprints on the exposed balloon surface, indicating that the deformed struts were initially crimped on the balloon.The balloon is well folded and shows no signs of inflation, but contrast medium residue was observed in the inflation lumen which indicates application of negative pressure.The angiographic material shows the implantation a stent in the target vessel.Thereafter, an unsuccessful attempt of crossing the proximal portion of the previously implanted stent with the complaint instrument is visible.After withdrawal of the complaint instrument, a shorter stent is successfully implanted.It remains unclear if the complaint stent was deformed prior to or during the crossing attempt.The angiographic material does therefore not provide any further relevant information with regards to the root cause of the complaint.Review of the product release documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined number of samples is tested from every lot.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause could be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 2.25/18
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key13167960
Report Number1028232-2022-00034
Device Sequence Number14786072
Product Code NIQ
UDI-Device Identifier07640130439023
UDI-Public7640130439023
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/11/2023
Device Model Number364487
Device Catalogue NumberSEE MODEL NO.
Device Lot Number04211310
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2022
Supplement Date Received by Manufacturer01/24/2022
Initial Report FDA Received Date01/05/2022
Supplement Report FDA Received Date01/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-