| Model Number |
CD2311-36Q |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This report is to advise of an event observed during analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
No complaint was received with the return of the device.Failure event observed during analysis.Final analysis found that the implantable cardioverter defibrillator exhibited premature battery depletion.Electrical testing revealed that there was a high current drain on the hybrid integrated circuit.The cause of the premature battery depletion was attributed to the component failure of the hybrid that led to high current drain.
|
| |
|