Model Number N/A
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Fall (1848); Necrosis (1971); Joint Dislocation (2374)
Date of Event 11/19/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant products: 010000980 3929945 g7 freedom const e1 lnr 32mm b; 110010261 6952447 g7 osseoti multihole 46mm b; competitor polar femoral component.Customer has indicated that the product will not be returned to zimmer biomet for investigation, due to the device was requested, but not returned by the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2021 - 03272.
 
Event or Problem Description
It was reported patient underwent right hemiarthroplasty to repair a femoral neck fracture.Subsequently 3 dislocations occurred resulting in conversion to right total hip arthroplasty approximately 3 moths post right hemiarthroplasty.Patient sustained a fall approximately 6 weeks post op that resulted in a dislocation and 2 additional dislocations occurred post reduction.Revision performed approximately 3 months post tha with removal of head/liner, and competitor stem.Surgeon noted necrosis of capsule and well-fixed acetabulum.Attempts have been made and additional information on the reported event is unavailable.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was confirmed by review of medical records received.Review of the available records identified findings of recurrent dislocations.Patient dislocated multiple times post-op after a previous revision.Necrosis, well-fixed cup, and previous cables were removed, no complications during this revision.The device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined, however, it was noted a competitor stem was used with a zimmer biomet head/liner, and while this is considered off-label use it is unknown if this caused or contributed to the event.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTRAINED HEAD
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13168617
Report Number0001822565-2022-00023
Device Sequence Number5422116
Product Code KWY
UDI-Device Identifier00889024499041
UDI-Public(01)00889024499041(17)300630(10)3036163
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K183457
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number802403203
Device Lot Number3036163
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/31/2021
Supplement Date Received by Manufacturer03/22/2022
Initial Report FDA Received Date01/05/2022
Supplement Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
Patient Weight48 KG
-
-