Catalog Number 66802001 |
Device Problems
Device Difficult to Setup or Prepare (1487); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Event Description
|
It was reported that, before wound therapy, it was observed that a pico 7 multisite large 20cm x 25cm dressing was already glued to the protective package.Bandages have never been applied to a patient¿s wound and treatment was completed with a smith & nephew back-up device.No patient injury reported.
|
|
Manufacturer Narrative
|
Reassessment of this incident found it not to fulfill reporting criteria.A device that is difficult to setup or prepare would not be likely to cause or contribute to a death or serious injury.Event may result in minor or temporary injury (e.G., pain) requiring no or minor medical intervention.The event may require use of a backup device to continue therapy.This event did not lead to a death or serious injury, nor would it be likely to cause or contribute to a death or serious injury if it were to recur.This event is considered not reportable pursuant to 21 cfr §803.Corrected h6.
|
|