• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE FIXED BEARING POSTERIOR STABILIZED (PS) LEFT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE FIXED BEARING POSTERIOR STABILIZED (PS) LEFT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during implantation, a loose sliver of an object was retrieved.The surgeon remarked that this was adhered to the articular surface, and had to be removed from the surgical site with the use of an artery forceps.It was remarked by the surgical team that this could have been part of the articular surface (as a manufacturing defect), and may not have been part of the cement used.The cement used during the surgery was either palacos mv or smartset ghv.It was reported that no further information is available.
 
Additional Manufacturer Narrative
(b)(4).Report source: foreign - (b)(6).The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The product was not returned for evaluation.Visual evaluation of the debris returned found the debris fragment to be visually consistent with the packaging foam insert used during manufacturing.However, as the debris was found after the sterile packaging was opened, the origin of the debris cannot be confirmed or conclusively linked to the manufacturing process or transit damage.The dhr was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause or condition of the device when it left zimmer biomet cannot be determined, as the product was opened during surgery.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g6; h1; h2; h3; h6; h10.The product was not returned for evaluation.Visual evaluation of the debris returned found the debris fragment to be visually consistent with the packaging foam insert used during manufacturing.The device history record was reviewed and no discrepancies relevant to the reported event were found.The condition of the device when it left zimmer biomet is conforming to specification.The root cause of the reported event can be attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE FIXED BEARING POSTERIOR STABILIZED (PS) LEFT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13169997
Report Number3007963827-2022-00002
Device Sequence Number13942052
Product Code MBH
UDI-Device Identifier00889024237254
UDI-Public(01)00889024237254(17)250228(10)64792823
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K121771
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model NumberN/A
Device Catalogue Number42512400711
Device Lot Number64680416
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/15/2021
Supplement Date Received by Manufacturer02/03/2022
04/26/2022
Initial Report FDA Received Date01/05/2022
Supplement Report FDA Received Date02/04/2022
04/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient Age63 YR
Patient SexMale
-
-