ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY EXTENDED VISION UV ABSORBING IOL; INTRAOCULAR LENS Back to Search Results
Model Number DAT015
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Disturbances (2140)
Date of Event 09/03/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints in the lot.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-healthcare professional reported that following an intraocular lens (iol) implant procedure, the patient had visual disturbances.The patient continued blurred/fluctuating vision and difficulty with glare at night.The patient could not adapt to the lens.The iol was exchanged with an alternate iol approximately 6 months following the initial procedure.There was no patient harm.
 
Additional Manufacturer Narrative
The explanted lens was not returned.An empty non-company lens case (non company lens diopter) was returned, removed from its pouch inside the opened non- company lens carton.Information was provided that indicated the use of a qualified cartridge and handpiece.Information was provided that indicated the use of two non-qualified viscoelastics.The root cause was determined to be unrelated to the lens.The explanted lens was not returned.A facility reported an explant due to visual disturbances and replaced with monofocal intraocular lens (iol).Surgeon reported iol exchange required due to patient could not properly adapt to the edof (extended depth of focus) lens.The replacement lens was reported to be a non-company monofocal model of the same diopter (non company lens diopter).The empty replacement lens case was returned.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY EXTENDED VISION UV ABSORBING IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13170048
Report Number1119421-2022-00028
Device Sequence Number1956060
Product Code POE
UDI-Device Identifier00380652424660
UDI-Public00380652424660
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/10/2025
Device Model NumberDAT015
Device Lot Number15098215
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/14/2021
Supplement Date Received by Manufacturer01/04/2022
Initial Report FDA Received Date01/05/2022
Supplement Report FDA Received Date01/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AMVISC.; HEALON.; MONARCH III CARTRIDGE D.; MONARCH III INJECTOR.
Outcome Attributed to Adverse Event Required Intervention; Required Intervention;
Patient SexUnknown
-
-