BOSTON SCIENTIFIC CORPORATION MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7580
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age at time of event: 18 years or older.
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in the left anterior descending artery.A 2.75mm x 12mm maverick 2 balloon catheter was selected for use.However, while advancing the balloon inside the guide catheter, the shaft broke.The procedure was not completed due to this event.There were no patient complications nor injuries reported and the patient status was stable.
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in the left anterior descending artery.A 2.75mm x 12mm maverick 2 balloon catheter was selected for use.However, while advancing the balloon inside the guide catheter, the shaft broke.The procedure was not completed due to this event.There were no patient complications nor injuries reported and the patient status was stable.
 
Additional Manufacturer Narrative
A2 age at time of event: 18 years or older.Returned product consisted of a maverick 2 mr balloon catheter.The device was visually and microscopically examined.At 19cm from the strain relief the hypotube was separated.The separated ends were ovaled in shape indicating the device was kinked prior to separation.The hypotube of the device has numerous kinks.There was contrast in the inflation lumen and the balloon was tightly folded.The shipping mandrel was still in place but was able to be removed without issue or resistance.The tip of the device was damaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13170093
Report Number2134265-2022-00028
Device Sequence Number1830097
Product Code LOX
UDI-Device Identifier08714729369998
UDI-Public08714729369998
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/12/2022
Device Model Number7580
Device Catalogue Number7580
Device Lot Number0024765549
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/16/2021
Supplement Date Received by Manufacturer03/07/2022
Initial Report FDA Received Date01/05/2022
Supplement Report FDA Received Date03/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/13/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-