ABBOTT POINT OF CARE I-STAT ACT KAOLIN CARTRIDGE Back to Search Results
Catalog Number 03P87-25
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Myocardial Infarction (1969)
Date of Event 12/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On (b)(6) 2021, abbott point of care was contacted by a customer regarding act kaolin cartridges that yielded multiple code 26 on a (b)(6) male patient with acute st elevation myocardial infarction (stemi) due to occlusion of circumflex coronary artery.This patient's procedure was not completed and the patient was transferred to another facility.Customer states there was heavy calcification was noted in the rca lesion with residual 98% lesion.Unable to cross the calcified lesion with a bigger diameter balloon.There was no additional patient information available at the time of this report.Return product is available for investigation.(b)(6).At this time there is no reason to suspect a malfunction exists.An investigation is underway.
 
Additional Manufacturer Narrative
Apoc incident: (b)(4).The investigation was completed on 18-jan-2022.A review of the device history record (dhr) confirmed the cartridge lot passed release specifications.Retained and returned cartridge testing met the acceptance criteria outlined in appendix 1 of q04.01.003 rev.Ag (product complaint level 2 and level 3 investigation procedure).No deficiency has been identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT ACT KAOLIN CARTRIDGE
Common Device Name
ACT KAOLIN CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
ottawa, ontario K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key13170143
Report Number2245578-2022-00001
Device Sequence Number11203718
Product Code JBP
UDI-Device Identifier10054749000156
UDI-Public10054749000156
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K023582
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/29/2022
Device Catalogue Number03P87-25
Device Lot NumberR21273
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/15/2021
Supplement Date Received by Manufacturer01/19/2022
Initial Report FDA Received Date01/05/2022
Supplement Report FDA Received Date01/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
-
-