| Model Number |
B9837 |
| Medical Device Problem Code |
Free or Unrestricted Flow (2945)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is pending investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported that a 102" (259 cm)appx 15.3ml 20 drop admin w/microclave®, 0.2 micron filter, flow controller, rotating luer was defective.It was further described that the dial-a-flow isn¿t working correctly and the flow rate is too fast.There was patient involvement but there was no harm or delay in therapy.
|
| |
|
Additional Manufacturer Narrative
|
|
One used 102" (259 cm)appx 15.3ml 20 drop admin w/microclave¿, 0.2 micron filter, flow controller, rotating luer was received and visually inspected.As received, no visible damage or anomalies were observed.No mating devices were returned.The set was primed while the flow regulator was in the open position.Unrestricted flow was established.A flow controller test was conducted and the product met performance specifications.The reported complaint could not be confirmed.The lot history was reviewed and no nonconformities were identified that may have contributed to the reported complaint.
|
| |
|