| Model Number |
BC0411XC-ID-700 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was discarded and not returned to the manufacturer for analysis; therefore, the alleged product issue cannot be confirmed.The instruction for use (ifu) include the warnings: do not exceed the maximum recommended inflation volume as balloon rupture may occur.Excessive pressure higher than 300 psi (2069kpa) may cause leakage or rupture of the balloon catheter working lumen.
|
| |
|
Event or Problem Description
|
|
It was reported that during treatment using a scepter xc balloon, the balloon burst using 70:30 contrast mix upon deployment of a guidewire.No complications or adverse clinical consequence were reported for the patient.
|
| |
|