• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PRECISION IDEAL EYES 10MM 45°, HD AUTOCLAVABLE LAPAROSCOPE 33CM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PRECISION IDEAL EYES 10MM 45°, HD AUTOCLAVABLE LAPAROSCOPE 33CM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 0502103045
Medical Device Problem Code Optical Distortion (3000)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that there was blurry image.
 
Event or Problem Description
It was reported that there was blurry image.
 
Additional Manufacturer Narrative
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: poor visual field.Probable root cause: incorrectly assembled optical train, damage to optical train, end of life wear-out, shipping damage, use error.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
 
Event or Problem Description
It was reported that there was blurry image.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION IDEAL EYES 10MM 45°, HD AUTOCLAVABLE LAPAROSCOPE 33CM
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key13170838
Report Number0002936485-2022-00004
Device Sequence Number17239559
Product Code GCJ
UDI-Device Identifier07613327062144
UDI-Public07613327062144
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K993045
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0502103045
Device Catalogue Number0502103045
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/08/2021
Supplement Date Received by Manufacturer12/08/2021
Initial Report FDA Received Date01/05/2022
Supplement Report FDA Received Date03/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-