| Model Number |
0502103045 |
| Medical Device Problem Code |
Optical Distortion (3000)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was blurry image.
|
| |
|
Event or Problem Description
|
|
It was reported that there was blurry image.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: poor visual field.Probable root cause: incorrectly assembled optical train damage to optical train end of life wear-out shipping damage use error the reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.H3 other text : 81.
|
| |
|
Search Alerts/Recalls
|