MAUDE Adverse Event Report: DEPUY MITEK LLC US ARTHRO SUTURE SCISSORS *EA; SUTURE CUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US ARTHRO SUTURE SCISSORS *EA; SUTURE CUTTER Back to Search Results
Model Number 214605
Medical Device Problem Codes Use of Device Problem (1670); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Udi: (b)(4).Incomplete.The lot number was unknown.Investigation summary: according to the information provided, it was reported that the mitek arthroscopic scissors are damaged.The complaint device is not being returned, therefore unavailable for a physical evaluation, however a photo was provided.Upon visual inspection of the photo, it was found that the device's internal shaft is bent.Given that lot number was not provided, the manufacturing record evaluation (mre) review cannot be performed.If the lot number becomes available, the mre review will be performed.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection of the photo, this complaint can be confirmed.The possible root cause can be attributed to hard and continuous use, since this is a reusable device, the constant manipulation between surgeries and sterilization process can lead to a damaged inner shaft.As per ifu; end of useful instrument life is generally determined by wear or damage from handling or surgical use.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Event or Problem Description
It was reported by the customer in (b)(6) that during an unknown surgery on (b)(6) 2021, it was observed that the arthro suture scissors device were damaged.During in-house engineering evaluation, it was determined that the internal shaft on the device was bent.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device is not being returned, therefore unavailable for a physical evaluation, however a photo was provided.Upon visual inspection of the photo, it was found that the device's internal shaft is bent.Given that lot number was not provided, the manufacturing record evaluation (mre) review cannot be performed.If the lot number becomes available, the mre review will be performed.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection of the photo, this complaint can be confirmed.The possible root cause can be attributed to hard and continuous use, since this is a reusable device, the constant manipulation between surgeries and sterilization process can lead to a damaged inner shaft.As per ifu 107066; end of useful instrument life is generally determined by wear or damage from handling or surgical use.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.The complaint device was received and evaluated.Upon visual inspection, it was found that the device's jaw is jammed, therefore, the internal shaft is bending while the handle is opened.At magnification of the jaw, damages were found on the device's cutting edge.A manufacturing record evaluation was performed for the finished device and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection of the photo, this complaint can be confirmed.The possible root cause can be attributed to hard and continuous use, since this is a reusable device, the constant manipulation between surgeries and sterilization process can lead to damages on the device's cutting edge.As per ifu 107066; end of useful instrument life is generally determined by wear or damage from handling or surgical use.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SUTURE CUTTER
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key13171571
Report Number1221934-2022-00019
Device Sequence Number11917240
Product Code FZT
UDI-Device Identifier10886705003884
UDI-Public10886705003884
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number214605
Device Catalogue Number214605
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/27/2021
Supplement Date Received by Manufacturer02/11/2022
02/21/2022
Initial Report FDA Received Date01/05/2022
Supplement Report FDA Received Date02/15/2022
02/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-