Model Number N/A
Medical Device Problem Codes Positioning Failure (1158); Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an articular surface was not seating to the tibial tray.There is no additional information available.
 
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.No product was returned, or pictures provided; therefore, visual and dimensional evaluations could not be performed.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.Visual examination of the returned product identified signs of use.The dovetail feature was flared and compressed.Damage to the articular surface can occur if the articular surface is not correctly placed and oriented before pushing it using the inserter instrument.Detailed instructions for inserting the articular surface are provided in persona- the personalized knee surgical technique (97-5026-001-00, rev.13) pg.44.The root cause is attributed to use error.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE MEDIAL CONGRUENT (MC) LEFT 14 MM HEIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13172127
Report Number3007963827-2022-00005
Device Sequence Number4517370
Product Code MBH
UDI-Device Identifier00889024469259
UDI-Public(01)00889024469259(17)260228(10)64966312
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K150090
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model NumberN/A
Device Catalogue Number42512100714
Device Lot Number64966312
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2021
Supplement Date Received by Manufacturer02/08/2022
06/27/2022
Initial Report FDA Received Date01/05/2022
Supplement Report FDA Received Date02/15/2022
06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
-
-