| Catalog Number |
320617 |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that 3 pn relion 31g x 6mm 3b tw were unable to deliver insulin or medication.The following information was provided by the initial reporter: the consumer reported that 3 pen needles clogged during injection.Date of event: (b)(6) 2021; samples status: discarded.
|
| |
|
Event or Problem Description
|
|
It was reported that 3 pn relion 31g x 6mm 3b tw were unable to deliver insulin or medication.The following information was provided by the initial reporter: the consumer reported that 3 pen needles clogged during injection.Date of event: (b)(6) 2021.Samples status: discarded.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Clog test was conducted on 7pcs retention samples and no needle clog fail found.
|
| |
|
Search Alerts/Recalls
|