MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU PN RELION 31G X 6MM 3B TW; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU PN RELION 31G X 6MM 3B TW; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Catalog Number 320617
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/05/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that 3 pn relion 31g x 6mm 3b tw were unable to deliver insulin or medication.The following information was provided by the initial reporter: the consumer reported that 3 pen needles clogged during injection.Date of event: (b)(6) 2021; samples status: discarded.
 
Event or Problem Description
It was reported that 3 pn relion 31g x 6mm 3b tw were unable to deliver insulin or medication.The following information was provided by the initial reporter: the consumer reported that 3 pen needles clogged during injection.Date of event: (b)(6) 2021.Samples status: discarded.
 
Additional Manufacturer Narrative
H6: investigation: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Clog test was conducted on 7pcs retention samples and no needle clog fail found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PN RELION 31G X 6MM 3B TW
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
no.1 liangpu street
suzhou industrial park
suzhou 21512 6
CH  215126
Manufacturer (Section G)
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
no.1 liangpu street
suzhou industrial park
suzhou 21512 6
CH   215126
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13172307
Report Number3014704491-2021-00415
Device Sequence Number4845147
Product Code FMI
UDI-Device Identifier00681131386807
UDI-Public00681131386807
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K162516
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/30/2025
Device Catalogue Number320617
Device Lot Number0248548
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2021
Supplement Date Received by Manufacturer01/07/2022
Initial Report FDA Received Date01/05/2022
Supplement Report FDA Received Date01/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-