| Model Number |
168-420 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
12/08/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation and exchange from textured to smooth.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported, left side deflation.And exchange from textured to smooth.Device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation and exchange from textured to smooth.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: opening, yellow particles/mold black in the surface of the shell.Leak test was performed and found opening on posterior side.A microscopic analysis was performed which identify opening sharp in the posterior side.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a sharp opening on posterior side to an unidentified (tear) opening.Non-penetrating nicks.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation and exchange from textured to smooth.Device has been explanted and replaced.
|
| |
|