Model Number 245775
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/10/2021
Type of Reportable Event Serious Injury
Event or Problem Description
The customer reports a patient had an adenoidectomy and right ear duravent tube placement , no issues reported.One year, and 17 days later, the patient required a myringotomy and tympanostomy (m+t) procedure and foreign body retrieval (of the duravent ear tube) due to migration of the duravent into the middle ear.There were no additional consequences to the patient reported.
 
Additional Manufacturer Narrative
The device referenced in this report was not returned to olympus for evaluation.The definitive cause of the user's experience cannot be determined at this time.The investigation is ongoing.This report will be updated upon completion of the investigation or upon receipt of additional relevant information.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The investigation does not suggest product non-conformances during manufacturing processes.Based on the results of the investigation, a possible root cause is the use error when trimming the duravent tube for the procedure.If the incision is made too large, the device may migrate into the middle year.However, the definitive root cause of the reported issue could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURAVENT VT TB 1.27MM TP (60/PK)
Common Device Name
TUBE, TYMPANOSTOMY
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
Manufacturer Contact
kenneth pittman
800 west park drive
westborough, MA 01581
9013785969
MDR Report Key13172828
Report Number3003790304-2022-00003
Device Sequence Number4780677
Product Code ETD
UDI-Device Identifier00821925018143
UDI-Public00821925018143
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
PREAMENDMENT
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number245775
Device Lot NumberNC799934
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/10/2021
Supplement Date Received by Manufacturer01/19/2022
Initial Report FDA Received Date01/05/2022
Supplement Report FDA Received Date02/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/30/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Required Intervention;
Patient Age4 YR
Patient SexFemale
-
-