| Model Number |
HS-004 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Infutronix is waiting for the iv set to be returned for evaluation.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021, infutronix received a complaint from an end user: "an administration set model hs-004 lot 2106005 was leaking at the connection below the filter." device operator was a patient.Medication infused was 5fu.A patient was involved but not harmed.The contract manufacturer of the affected device (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A service provider of infutronix provided the evaluation report for the affected administration set on (b)(6) 2022.Visual inspection confirmed: there was presence of dried 5fu medication on the band-aid wrapped around the tubing above the filter, indicating leakage of the medication.The reported leaking issue was confirmed.
|
| |
|
Event or Problem Description
|
|
This is a follow-up for the initially filed mdr (b)(4).
|
| |
|
Search Alerts/Recalls
|