MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BARE PL 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BARE PL 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number APB-1.5-4-3D-ES
Medical Device Problem Code Stretched (1601)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2021
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received a report that after the normal angiography was over, the access catheter was established.The first coil was pushed in the micro-catheter and continued to fill, and the second coil was in the process of filling but it could be seen under the image that it was stretched in the body.The coil was replaced and refilled, and the operation proceeded smoothly.There were no patient symptoms or complications associated with the event.The device and accessory devices were prepared as indicated in the ifu.The patient was undergoing surgery for treatment of a saccular, unruptured aneurysm in the internal carotid posterior communicating artery with a max diameter of 4mm and a 2mm neck diameter.It was noted the patient's blood flow was normal and vessel tortuosity was normal.
 
Additional Manufacturer Narrative
The axium prime coil and was returned for analysis within a shipping box; within a sealed plastic biohazard pouch and within a dispenser track.The unknown catheter used during the event was not returned.Visual inspection/damage location details: the axium implant coil was returned within the introducer sheath.The implant coil was found to be inverted within the introducer sheath.The coil was found to be located in the ¿wave-lock¿ portion of the introducer sheath.The axium implant coil pushwire was found to be bent at ~77.8cm from proximal end.The axium implant coil was found to be broken within introducer sheath.The axium implant coil was found to be missing.Conclusion: based on the device analysis and reported information, the customer¿s report of ¿coil stretch¿ was unable to be confirmed as the implant coil was found to be broken and missing.The implant coil was found to be inverted within the introducer sheath.It is possible that the customer inadvertently inverted the axium detachable coil after inspection or hydration, following the event, or during shipping back to medtronic for evaluation.There was no reported issue pushing the axium prime coil out from within the introducer sheath.Therefore, it is possible the axium coil became damaged upon removal from within the introducer sheath; however, the root cause could not be determined.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
H3: product analysis #704682258:equipment used: vis (m-78210), 203cm ruler (m-83361) document used: n/a as found condition (condition of returned device): the axium prime coil and was returned for analysis within a shipping box; within a sealed plastic biohazard pouch and within a dispenser track.The unknown catheter used during the event was not returned.Visual inspection/damage location details: the axium implant coil was returned within the introducer sheath.The implant coil was found to be inverted within the introducer sheath.The coil was found to be located in the ¿wave-lock¿ portion of the introducer sheath.The axium implant coil pushwire was found to be bent at ~77.8cm from proximal end.The axium implant coil was found to be broken within introducer sheath.The axium implant coil was found to be missing.Testing/analysis (including sem reports): n/a conclusion: based on the device analysis and reported information, the customer¿s report of ¿coil stretch¿ was unable to be confirmed as the implant coil was found to be broken and missing.The implant coil was found to be inverted within the introducer sheath.It is possible that the customer inadvertently inverted the axium detachable coil after inspection or hydration, following the event, or during shipping back to medtronic for evaluation.There was no reported issue pushing the axium prime coil out from within the introducer sheath.Therefore, it is possible the axium coil became damaged upon removal from within the introducer sheath; however, the root cause could not be determined.(reyesr32 2021-12-21) medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PRIME BARE PL 3D
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine,ca 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine,ca 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key13173427
Report Number2029214-2022-00006
Device Sequence Number4998332
Product Code KRD
UDI-Device Identifier00847536021167
UDI-Public00847536021167
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K151447
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/08/2024
Device Model NumberAPB-1.5-4-3D-ES
Device Catalogue NumberAPB-1.5-4-3D-ES
Device Lot Number222382268
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/21/2021
Supplement Date Received by Manufacturer10/09/2021
Initial Report FDA Received Date01/05/2022
Supplement Report FDA Received Date11/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age54 YR
Patient SexMale
-
-