As the lot number for the device was provided, a review of the device history record is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a video was provided for review.The investigation of the reported event is currently underway.(expiry date: 07/2023).
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.One video was provided and reviewed.The video shows the health professionals inflating the conquest balloon with an inflation pump.Water is noted to be leaking from the distal end of the balloon.However the specific cause of the leak can not be identified in the video.No other specific anomalies and no evidence of balloon rupture noted.Therefore, based on the video review, the reported balloon rupture is kept inconclusive.A definitive root cause for the alleged balloon rupture could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 07/2023),.
|