Based on the information provided, review of the surgical technique manual, ifu, certificates of conformance and fmea, there is no indication of device failure and no indication that the devices were out of specification.The most probable root cause is user error: not installing the implants fully across the si joint.Additionally, per 300223-us-a, si-bone surgical technique manual (c-arm and o-arm), the surgeon is instructed as follows: "prior to closure, always obtain final fluoroscopic images in the lateral, inlet, and outlet views to confirm no cortical wall breach, foramen breach, or other malposition.".
|