• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAYER MEDICAL CARE, INC. SPECTRIS SOLARIS SYRINGE KIT; ANGIOGRAPHIC SYRINGE KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAYER MEDICAL CARE, INC. SPECTRIS SOLARIS SYRINGE KIT; ANGIOGRAPHIC SYRINGE KIT Back to Search Results
Model Number 60728591
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/21/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Bayer product analysis received and examined the returned mr disposable syringe kit.Visual examination of the syringe barrel confirmed the presence of a white particle within the fluid path that was most likely material from the inner wall of the contrast spike luer which occurred during the component manufacturing/assembly process.A 12 month review of complaint history determined the complaint rate to be (b)(6) complaints per (b)(6).
 
Event or Problem Description
The customer reported the following: a white foreign body was found inside the syringe upon opening a new ssqk65/115vs syringe kit.No injury or adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRIS SOLARIS SYRINGE KIT
Common Device Name
ANGIOGRAPHIC SYRINGE KIT
Manufacturer (Section D)
BAYER MEDICAL CARE, INC.
1 bayer drive
indianola PA 15051
Manufacturer (Section G)
BAYER MEDICAL CARE, INC.
1 bayer drive
indianola PA 15051
Manufacturer Contact
susan sherwin
1 bayer drive
indianola, PA 15051
7249408678
MDR Report Key13175399
Report Number2520313-2022-00002
Device Sequence Number11917273
Product Code DXT
UDI-Device Identifier00616258007786
UDI-Public(01)00616258007786
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K012950
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/07/2026
Device Model Number60728591
Device Catalogue NumberSSQK65/115VS
Device Lot Number8417225
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2021
Initial Report FDA Received Date01/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/07/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-