Model Number FV410T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 09/12/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This complaint was reopened in reference to (b)(4).Manufacturing site evaluation: manufacturing and quality control data the progav® was manufactured by a qualified employee in july 2015.Deviations during assembly did not occur.The valve was sterilized by miethke and released for shipment after final inspection.The progav® has a normal pressure range of 0 to 20 cmh2o.The shunt system was inspected as article (b)(4).All parameters have been inspected and approved during the manufacturing process.All parameters have been assessed as meeting specifications.Device examination: the progav® was returned submersed in an unidentified liquid.Visual inspection: scratches on the outer housing of the valve were observed through the visual inspection.However, no significant deformations or damage were detected.Permeability test: a permeability test has shown that the valve had a blockage.Adjustment test: the progav® was tested and was adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function was fully operational and the braking force was within expected tolerances.Computer controlled test: a computer controlled test was not able to be performed because the valve was not penetrable.Result: first, a visual inspection of the progav® was performed.Scratches on the outer housing of the valve were observed through the visual inspection, however, no significant deformations or damage were detected.Next, the permeability, adjustability, as well as the brake functionally and brake force, of the valve were tested.The valve was not permeable, therefore, a computer controlled test was not able to be performed.All other specifications were met.Finally, the valve was dismantled.A build-up of a substance (likely protein) was observed inside the valve.Based on the investigation, a valve occlusion was able to be confirmed which was likely due to the deposits observed inside the valve.Deposits caused by natural substances in the cerebrospinal fluid, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Manufacturing defects at the time of release have been excluded.The valve met all specifications of the final inspection when released from christoph miethke (b)(4) no further regulatory actions are required.
 
Event or Problem Description
It was reported to miethke that a progav valve (part # fv410t) was implanted during a procedure performed on (b)(6) 2015.According to the complainant, the valve was suspected of operating in over-drainage.The patient underwent a revision procedure on (b)(6) 2017.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: (b)(6), weight: (b)(6) (kg), height: 160 centimeters (cm), gender: unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV VALVE
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key13175882
Report Number3004721439-2021-00825
Device Sequence Number19989698
Product Code JXG
UDI-Device Identifier04041906132071
UDI-Public4041906132071
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K103003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Model NumberFV410T
Device Catalogue NumberFV410T
Device Lot Number20027140
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2017
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2017
Initial Report FDA Received Date01/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/09/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age15 YR
Patient SexUnknown
Patient Weight50 KG
-
-