MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE FIXED BEARING CRUCIATE RETAINING (CR); PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE FIXED BEARING CRUCIATE RETAINING (CR); PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Ambulation or Postural Difficulties (2544)
Date of Event 05/03/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: 42502006002 - femur cemented cruciate retaining (cr) narrow right size 6 - 64759285.42532006702 - tibia cemented 5 degree stemmed right size d - 64831031.42540000032 - all poly patella cemented 32 mm diameter - 6483311.Customer has indicated that the product will not be returned to zimmer biomet for investigation as it remains implanted.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2022-00006.
 
Event or Problem Description
It was reported a patient underwent an initial right total knee arthroplasty.Subsequently, the patient reported severe pain and difficulties with limited range of motion and the patient underwent an uncomplicated manipulation under anesthesia approximately 6 weeks later.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.D10: unknown - unknown palacos cement - unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: post op: pain is controlled with narcotics, no complication on x-rays; mua performed: rom during initial procedure reached 0-130 degrees, knee reached 120 degrees with gentle force; follow up: severe pain, unable to do activities, neutral alignment, no laxity/instability, rom to 62 degrees, no complication on x-rays.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE FIXED BEARING CRUCIATE RETAINING (CR)
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13175900
Report Number3007963827-2022-00007
Device Sequence Number17306796
Product Code JWH
UDI-Device Identifier00889024241558
UDI-Public(01)00889024241558(17)250819(10)64766911
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/19/2025
Device Model NumberN/A
Device Catalogue Number42522000410
Device Lot Number64766911
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2021
Supplement Date Received by Manufacturer01/11/2022
04/14/2022
Initial Report FDA Received Date01/06/2022
Supplement Report FDA Received Date02/03/2022
04/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age46 YR
Patient SexFemale
Patient Weight87 KG
Patient RaceBlack Or African American
-
-