ABBOTT MEDICAL AMPLATZER VASCULAR PLUG II; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 9-AVP2-006
Medical Device Problem Codes Break (1069); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/22/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2021, a 6 mm amplatzer vascular plug ii was selected for implant.During the procedure, the physician was unable to pass the device through a 5f non-abbott catheter.The physician noted that the amplatzer vascular plug ii was damaged, so the device was exchanged for a new device, which was successfully implanted to resolve the event.The patient remained hemodynamically stable throughout the procedure, however there was a clinically significant delay.The patient is stable.
 
Additional Manufacturer Narrative
The investigation results will be provided in the final report.
 
Additional Manufacturer Narrative
An event of difficulty inserting the device into a competitor catheter and the device being damaged could not be confirmed.The investigation confirmed the device met dimensional, visual, and functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER VASCULAR PLUG II
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key13175901
Report Number2135147-2021-00591
Device Sequence Number11925755
Product Code KRD
UDI-Device Identifier00811806010366
UDI-Public00811806010366
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K071125
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Model Number9-AVP2-006
Device Catalogue Number9-AVP2-006
Device Lot Number8025158
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2021
Supplement Date Received by Manufacturer02/02/2022
Initial Report FDA Received Date01/06/2022
Supplement Report FDA Received Date02/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
-
-