|
H10: device evaluation: two (2) sample were received in used conditions.Functional test: air cuff symmetry test was performed to the units according to mp bivona tt-tj130 rev.102 "leak test" and av-10006853 rev.101 - leak testing and cuff symmetry visual aid.Cuff symmetry fixture was used (id: (b)(6) due: 10/2022).Results: when the sample was inflated with air, the cuff only inflated one side.According to the ifu (10018853-001 rev.100, instructions for use - pedi/neo tts ifu (pnt)) the pilot balloon was manipulated, and the cuff inflated.After the whole cuff inflated, it was deflated and inflated 4 times, all the times the cuff inflated completely.When measuring the cuffs, the percentage for the symmetry were: sample 1= 35%; sample 2= 40%.These results are over 33.3% that is the minimum acceptable.Based on the test, the units are within spec.The complaint is not confirmed.No root cause could be determinate since the samples successfully passed cuff symmetry test.
|